Hach HIAC 9703+ Manuel D'utilisation page 18

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 21
1. Put a sample container on the module. If a small vial is used, refer to
Put a vial in the small-vial clamp
2. Push the SET button to lower the sample probe to the correct
position.
3. Push the LIFT/LOWER button to raise the sample probe to the top
position.
4. Push the LIFT/LOWER button to automatically move the sample
probe down and stop at the correct bottom position.
Note: This position is recognized by the external software.
The bottom position is saved until a new bottom position is set.
Flush the syringe
Flush the syringe when a new syringe is installed or when a new sample
is used. To optimize the performance of the syringe:
• Do not operate the syringe for more than a few cycles without liquid.
• Thoroughly flush the syringe with distilled or deionized water after use.
1. Push the LIFT/LOWER button to raise the sample probe.
2. Put the container with the flushing liquid under the sample probe.
3. Push the LIFT/LOWER button to lower the sample probe in the
flushing liquid.
4. Push the CLEAN button. The syringe fills with the flushing liquid and
pushes the liquid out to the drain.
Note: The system can also be flushed with the external software.
Put a vial in the small-vial clamp
The small-vial clamp holds small vials under the center of the sample
probe. To install the clamp assembly, refer to
on page 14.
18 English
on page 18.
Install the small-vial clamp
Stop all operations
The STOP button can be used to simultaneously stop movement of the
syringe and sample probe. Use the buttons or external software to
resume operations.
Connect the computer cable
A USB cable connects the instrument to a computer.
1. Install a compatible software application on the computer.
2. Turn on the instrument power.
3. Connect the USB cable to the USB port on the instrument and to the
computer.
Operate the instrument by computer
The instrument must connect to a computer for configuration, calibration,
data collection and data management. Refer to the software user
manual for specific instructions.

Publicité

Table des Matières
loading

Table des Matières