Electromagnetic Immunity; Electromagnetic Immunity / Mobile Radiofrequency Equipment - Satelec Acteon P5 Newtron Manuel D'utilisation

Masquer les pouces Voir aussi pour P5 Newtron:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 20

8. 2 ELECTROMAGNETIC IMMUNITY

The device is intended for use in the electromagnetic environment specified in the table below.
The user and/or installer must ensure that the device is used in such an electromagnetic environment.
Immunity test
Electrostatic
± 6KV contact
discharge (ESD)
± 8KV air
IEC 61000-4-2.
Electrical fast
transients
± 2KV for power supply lines
IEC 61000-4-4.
Surges
± 1KV differential mode
IEC 61000-4-5.
± 2KV common mode
<5% U
cycles.
40% U
Voltage dips, short
cycles
interruptions and
voltage variations
70% U
IEC 61000-4-11.
cycles
<5% U
cycles

8. 3 ELECTROMAGNETIC IMMUNITY / MOBILE RADIOFREQUENCY EQUIPMENT

The device is intended for use in the electromagnetic environment specified in the table below.
The user and/or installer must ensure that the device is used in such an electromagnetic environment.
Immunity test
Portable and mobile radiofrequency communications devices must not be used near the device (including its cables) at a distance less than that
Conducted
disturbance,
radiofrequency fields.
IEC61000-4-6
Radiated
radiofrequency
electromagnetic
field.
IEC61000-4-3
The electromagnetic field strengths of fixed radiofrequency emitters, as determined by an electromagnetic environment measurement (a),
must be less than the compliance level in each frequency range (b).
Interference may occur near equipment marked with the symbol below:
16
IEC 60601 test level
(>95% dip in U
) for 0.5
T
T
(60% dip in U
) for 5
T
T
(30% dip in U
) for 25
T
T
(>95% dip in U
) for 250
T
T
IEC 60601 test level
recommended and calculated according to the frequency and power of the emitter.
3 V/m
150 KHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
Compliance level
± 6KV contact
± 8KV air
± 2KV for power supply lines
± 1KV differential mode
± 2KV common mode
<5% U
(>95% dip in U
) for 0.5
T
T
cycles.
40% U
(60% dip in U
) for 5
T
T
cycles
70% U
(30% dip in U
) for 25
T
T
cycles
<5% U
(>95% dip in U
) for 250
T
T
cycles
Compliance level
Recommended separation distance:
3 V/m
d = 1.2
d = 1.2
d = 2.3
3 V/m
Where P is the maximum power rating of the emitter in watts
(W) according to the manufacturer's specifications and d is the
recommended minimum separation distance in meters (m).
Electromagnetic environment - guidance
Floors must be wood, concrete, cement or tiled.
If floors are covered with synthetic material
(carpet, etc.), the relative humidity must be at
least 30%.
Power quality should be that of a typical
commercial or hospital environment (hospital,
clinic).
Power quality should be that of a typical
commercial or hospital environment.
Power quality should be that of a typical
commercial or hospital environment.
If the use of the device requires continued
operation during a power cut, it is recommended
that the product be powered from a separate
power supply (UPS, etc.).
Electromagnetic environment - guidance
80MHz à 800MHz.
800MHz à 2.5GHz.

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

P5 newtron led

Table des Matières