d) Drying
Dry with a clean, non-woven single-use cloth, so
as to remove any traces of liquid.
Keep in single-use sterilization sleeves or pouches
in compliance with the specifications defined in
standard EN ISO 11607-1 or in any equivalent
standard required by the national regulations.
e) Sterilization
The tips must be sterilized individually in an
autoclave,
according
parameters:
- autoclave: type B, compliant with standard EN
13060;
- sterilization temperature: 134°C;
- sterilization stabilization time: 18 minutes;
- pressure: 2 Bars min.
f) Storage
Then, store the sterilized products in a dry place
protected from dust.
Before re-use, if the integrity of the packaging is
not conformable, recondition the re-sterilize in
accordance with the defined protocol.
If contamination is visible inside the bag, destroy
the product.
g) Disposal of the product
Dispose of the product in receptacles for waste
materials of healthcare activities involving
infectious risks.
CAUTION: Not all autoclaves can reach 134°C.
Not all autoclaves draw a pre-vacuum. Please
refer to your autoclave manufacturer for
specific sterilization instructions.
12
to
the
following

6. 3 DEVICE MAINTENANCE

Important:
- Do not use the device if it appears to be
faulty.
- Do not use an abrasive cleaning agent on the
device.
- Do not use any cleaning agent in spray form
on the device.
a) Maintenance
The device control unit, the handpiece cord and
the footswitch must be cleaned and disinfected
on a daily basis (using alcohol or disinfectants,
dental practice wipes, such as S
b) Monitoring
The accessories and cords should be regularly
checked for faulty insulation. Replace them if
necessary.
It is important to keep the control unit
ventilation vents clean in order to avoid
abnormal heating.
Check that there is no humidity in the handpiece
connector. If there is, dry it out by wiping, then
blowing with the multifunction syringe.
Note:
In the case of a fault, you are advised to contact
the supplier of your device. Do not use
unauthorized repairers, who might make your
device dangerous for you and your patients.