Satelec Acteon P5 Newtron Manuel D'utilisation page 14

Masquer les pouces Voir aussi pour P5 Newtron:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 20
F
AULT
No operation
No spray
No water at tip but
vibration present
Insufficient power
Low vibration
No ultrasonic output
No light
(P5 N
L
)
EWTRON
ED
Water leak between
handpiece and its
base or at junction
between handpiece or
cord
14
P
OSSIBLE CAUSE
Defective mains cord connection
Blown fuse
Defective water nose connection
No water pressure
Clogged filter or defective solenoid
valve
Obstructed tip or file
Wrong tip selection
Incorrect spray adjustment
Worn or deformed tip
Incorrect use : wrong angle of
application or inadequate pressure
on tooth
Presence of fluid or moisture
between handpiece and cord
Incorrect tip tightening
Defective connector contact
Cut wire in handpiece cord
Handpiece light ring is missing.
Faulty light ring.
Faulty light ring connector
contacts.
Light ring reversed polarity.
Faulty cord / handpiece connector
contacts.
Other.
Worn sealing gasket (115 x 1) in
handpiece
R
EMEDY
Check mains socket
Return to SATELEC After Sales Service
Return to SATELEC After Sales Service
Check water supply
Check pressure of mains
Clean or replace filter
Return to SATELEC After Sales Service
Remove obstruction for tips or file
Check tip
Adjust spray
Replace tip
See tip utilization method in T
Thoroughly dry electrical contacts
Tighten with wrench
Clean cord contacts
Return to SATELEC After Sales Service for
cord replacement
Install the light ring (Fig. 6).
Replace the light ring (Fig. 6).
Clean the connector light ring contacts.
Position the light ring by lining up the lug
+.
Clean the handpiece and/or connector
contacts.
Contact your installer-integrator.
Replace gasket (kit F12304)
IPBOOK

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

P5 newtron led

Table des Matières