Connection And Installation Of The Tm 235 Kf Titration Stand And Titration Vessel - Xylem TitroLine 7800 Mode D'emploi

Titrateur
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

159
2.6

Connection and installation of the TM 235 KF titration stand and titration vessel

Place the TM 235 KF titration stand to the right of the titrator and connect to the 12V out-socket in the rear panel
of the piston burette by using the TZ 1577 connection cable. The stand rod is screwed into the thread of the
TM 235 KF.
The titration vessel TZ 1770 is mounted at the stand rod. Please take care that the metal clamp is adjusted as
shown in the attached (Fig. 9).
Fig. 9
Put all three white inner plastic adapters at the waste, solvent and moisture bottle.
Fill the moisture bottle with molecular sieve. Connect the PVC and PTFE plastic tubes as shown in the next
pictures (Fig. 10 - Fig. 14):
The PVC tubes are connected to the connectors at the back side of the TM 235 KF.
The long PVC tube is used for the connection of the waste bottle.
The two shorter PVC ones are used to connect the moisture bottle and the solvent bottle.
Fig. 10

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Si analytics titroline 7000

Table des Matières