OHAUS a-AB33M1 Mode D'emploi page 44

Analyseur multiparamètre; ph-mètres de paillasse; conductimètre de paillasse
Masquer les pouces Voir aussi pour a-AB33M1:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

EN-42
Conductivity Meter Model
Measurement Channels
Measuring Range
Resolution
Conductivity
Accuracy
Reference
Temperature
Cell Constants
Temperature
Compensation
Measuring range
Resolution
TDS
Accuracy
TDS Factor Range
Measuring range
Resolution
Resistivity
Accuracy
Measuring range
Practical
Resolution
Salinity
Accuracy
Measuring Range
Resolution
Temperature
Accuracy
Calibration
Calibration points
Calibration
Calibration Sign
Calibration mode
Table 12-3 Specification 3
a-AB33EC
Conductivity/TDS/Salinity/Resistivity with
temperature
0.01 μS/cm to 19.99 μS/cm
20 μS/cm to 199.9 μS/cm
200 μS/cm to 1999 μS/cm
2.00 mS/cm to 19.99 mS/cm
20.0 mS/cm to 199.9 mS/cm
0.01 μS/cm minimum; auto ranging
± 0.5 % Reading ± 2 LSD
20 °C, 25 °C
0.01 to 10.00 cm
Linear (0 to 10.0%/ºC), off
0.1 mg/L to 200 g/L
0.01 mg/L minimum, auto ranging
± 0.5 % Reading ± 2 LSD
Linear, 0.04 to 10.00, default 0.5
1 to 100 MΩ-cm
0.01 Ω-cm auto ranging
± 0.5 % Reading ± 2 LSD
0 to 100 psu
0.01 psu minimum, auto ranging
± 0.5 % Reading ± 2 LSD
-5.0 to 110.0°C,
32.0 °F to 212.0 °F
0.1 °C, 0.1 °F
± 0.3 °C, ±0.3 °F
No
1 point cell constant calibration; 6 buffers
available (10, 84, 146.5, 500, 1413 μS/cm,
12.88mS/cm)
Cell constant & face
Linear
Analyzer and Bench Meter
-1

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

A-ab33phAb41phAb33ec

Table des Matières