19.5 Alibi memory

The weighing scale software can save up to 100 000 measurements. Measurements
are saved automatically each time the PRINT-key is pressed. In addition to
measurements other data is also saved:
• Date of measurement
• Time of measurement
• Measuring result (weight)
• Tare value
• Logged-in user
• Name of weighed product
• Measurement one will be deleted as soon as measurement 100 001 is
saved.
• Measurements saved to the alibi memory cannot be deleted.
Each can be invoked as described above (chap. 18.4).
If menu item b4.5.1.7 is selected and a printer connected, data can be printed by
pressing PRINT. (Dependent on settings selected in GLP report chap. 15).