ELECTRICAL SPECIFICATIONS

Measurement extent :
FUNCTION
V / m
µW / cm²
A / m
Pass band: from 100kHz to 2.5GHz
The measurement from 100kHz to 1MHz is purely indicative.
Specified measurement domain :
Measurements are taken in a distant field in order to obtain a flat wave.
The impedance of the ambient field must be equal to 377 W .
RANGE
Resolution
Accuracy (2)
Stability
(1) The power density is limited to the maximum display capacity of 1999µW/cm² corresponding to a field of
86.8V/m.
(2) Instrument only (without probe): ± 0.5% of the reading ± 0.2% of the range.
Error due to the interchangeability of sensors: ± 0.5dB.
Reference conditions
Distortion magnitudes
Ambient temperature
Relative humidity
Battery voltage
Field frequency
Field level

SPECIFICATIONS

V/m
V/m
from 0
from 1
to 1 V/m
to 10 V/m
0.1 V/m
0.1 V/m
0.7 V/m
0.5 V/m
Reference conditions
MEASUREMENT EXTENT
0.1 to 199.9
0.1 to 1999
0.1 to 19.99
V/m from V/m from
10 to
100 to
100 V/m
199.9 V/m
0.1 V/m
0.1 V/m
1 dB
2 dB
0.2 dB
20 °C
60 % RH
9 V
150 MHz
10.0 V/m
79
µW/cm²
from 0.1 to
from 200 to
199.9 µW/cm² 1999 µW/cm² (1)
0.1 µW/cm²
1 dB
Tolerances
± 2 K
± 10 %
± 1 V
± 1 %
± 0.1 V/m
µW/cm²
1 µW/cm²
2 dB