Self-Test Procedure

Power-On Self-Test

Each time the instrument is powered on, a self-test is performed. This test assures you that the
instrument is operational.
Self-test checks that the minimum set of logic and power mesh systems are functioning properly. Self-
test does not enable the output or place any voltages on the output. It leaves the instrument in the
reset
state.

User-Initiated Self-Test

The user-initiated self-test is the same as the power-on self-test.
Front Panel Menu Reference
Cycle ac power.
If self-test fails, the front panel ERR indicator comes
on. Press the Error key to display the list of errors.
For a list of errors, see
SCPI Error
304
SCPI Command
*TST?
If 0, self-test passed. If 1, self-test failed.
If self test-fails, use
self-test error.
Messages.
SYSTem:ERRor?