Programming The Output; Set The Output Priority Mode - Keysight Rp7900 Serie Guide D'utilisation Et D'entretien

Table des Matières
Les langues disponibles

Les langues disponibles

4 Using the Regenerative Power System

Programming the Output

Set the Output Priority Mode

Set the Output Voltage
Set the Output Current
Set the Slew Rate
Set the Output Resistance
Set the Output Bandwidth - RP795xA, RP796xA
Set the Output Bandwidth - RP793xA, RP794xA
Set the Output Turn-On/Turn-Off Mode - RP793xA, RP794xA
Enable the Output
SHOCK HAZARD, LETHAL VOLTAGES Many models generate voltages greater than
60 VDC, with some models rated at 950 VDC! Ensure that all instrument
connections, load wiring, and load connections are either insulated or covered using
the safety covers provided, so that no accidental contact with lethal voltages can
occur.
When the RPS is first turned on, it may take about 30 seconds or so to initialize the
instrument before it is ready for use.
Set the Output Priority Mode
Select either voltage or current priority mode. Refer to
Voltage priority mode keeps the output voltage constant. The output voltage remains at its
programmed setting, provided the load current remains within the + or - current limit setting.
Current priority mode keeps the output current constant. The output current remains at its
programmed setting, provided the load voltage remains within the voltage limit setting.
Front Panel Menu Reference
Select Output\Mode.
Select either Voltage priority or Current priority.
Then press Select.
When switching between voltage priority and current priority mode, the output is
turned off and the output settings revert to their Power-on or RST values.
112
Priority Mode Operation
SCPI Command
To specify current or voltage priority mode:
FUNC CURR|VOLT
Keysight RP7900 Series Operating and Service Guide
for more information.

Chapitres

Table des Matières
loading

Table des Matières