5 - OPERATOR INSTRUCTIONS

5.1 - CONTROLS

The appliance is equipped with the following control devices (Fig. 2):
- main power ON/OFF switch (1) with indications "1" (ON) and "0" (OFF);
- red AL (fault) indicator LED (4);
- yellow "CUT" button to select the cutting function, complete with yellow
indicator LED (2);
- blue "COAG" button to select the coagulation function, complete with
blue indicator LED (3);
- "POWER" control buttons "▲" - "▼" complete with display
(reading in W with tolerance of ± 20%) (5)
1
5.2 – INSTRUCTIONS FOR USE

5.2.1 - Setting into operation

After performing the installation procedures, the appliance can be set into
operation (Fig. 2):
1) fit the required electrode on the handpiece as indicated in § 4.1;
2) select the required function using the "CUT" (2) and/or "COAG" (3)
buttons;
3) switch on the appliance by means of the main power switch (1);
4) select the required power level by means of the "POWER" buttons "▲"
- "▼" (5);
5) instruct the patient to grip the neutral electrode;
6) press the button on the handpiece or press the pedal control; this
action will start the audible signal and deliver power to the electrode
attached to the handpiece.