Tecno-gaz ONYX B 6.0 Mode D'emploi page 90

Stérilisateur à vapeur

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 97
ENGLISH
 
Reading of the T2 probe during
A 653
sterilization above the maximum limit
A 661
Error in the probe reading
A 662
Error in the probe reading
Error for failed pressure achievement
A 701
during the first two phases of the
warm-up
Error for failed pressure achievement
A 711
during the third phase of the warm-up
Reading of the T1 probe during
A 751
sterilization below the minumum limit
Reading of the T2 probe during
A 753
sterilization below the minumum limit
Temperature of the 121°C cycle out o
A 781
the maximum limit
Temperature of the 134°C cycle out o
A 782
the maximum limit
Errors of out of maximum time during
A 801
the first phases of discharge
Errors of out of maximum time during
A 811
the last phase of discharge
Cycle interrupted due to power supply
A 901
failure
 
 
It's need to make a SD CARD periodic backup.
In the event of one of the alarms persisting after some time, consult technical assistance.
 
 
 
 
 
 
42
 
 Let cool the room
 Reset and restart the system
 Reset and restart the system
 Reset and restart the system
 Reset and restart the system
 Reset and restart the system
 Reset and restart the system
 Reset and restart the system
 Let cool the room
 Reset and restart the system
 Let cool the room
 Reset and restart the system
 Clean the filter in the room
 Reset and restart the system
 Clean the filter in the room
 Reset and restart the system
 Check the power supply system of
the machine and of the room.
 Reset and restart the system