10.9 Discharge used water
 
When the used water level LED (Fig. A-pos.
16
) lights up, the used water collection tank
must be emptied.
If nothing is done, the operation of the autoclave is blocked.
Take the hose supplied (Fig.8), and insert it into the used water drain fitting located on the
front of the autoclave (Fig.A-pos.
15
). Put the other end of the hose in a container and
unscrew the ring nut, turning it anticlockwise, the water will fall into the container, emptying
the tank.
IMPORTANT:
A - The hose located in the receiving container must never touch or be immersed in the
discharged water, otherwise sucking up will occur.
B - Always wait until the discharge water has been completely drained. The used water
maximum level LED turns off when some water still remains into the tank, consequently do
not considered it as a reference for this operation.
C - Any residues inside the drain tank are due to the normal functioning of the autoclave
and do not create problems for the sterilization cycles.
 
When drained, screw in the ring nut and remove the hose
10.10 Interruption of a sterilization cycle
 
A sterilization cycle can be voluntarily interrupted by pressing the button
Stop
for at least 2
seconds.
The autoclave will send out a beep and the display will show the alarm message A001.
For reset the alarm, press at the same time, the multifunction push buttons on the
Reset
bar up to the cancelling of the same.