Sartorius ACCULAB ALC Série Mode D'emploi page 14

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Example
Determine an unknown percentage; store the weight on the balance/scale as a reference percentage
Settings (changes in the factory settings required for this example):
Menu: Application program: Weighing in percent (menu code 2. 1. 5)
Menu: Reference percentage 100 % (menu code 3. 3. 5)
Step
1. Turn on the balance/scale
2. Tare the balance/scale
13. Display the reference
percentage
14. Place the reference weight for 100%
on the balance/scale (here: 222.5 g)
15. Start application; if the print
format is set to include data
ID codes, the following is printed:
16. Place an unknown weight
on the balance/scale (here: 322.5 g)
17. Display weight
18. Display percentage
19. Unload the balance/scale
10. Delete the reference percentage
11. Repeat the procedure starting from step 6, if desired.
14
Key (or instruction)
[ON/OFF]
[ZERO]
[F] > 2 sec.
[F]
[F]
[F]
[CAL] > 2 sec.
Display/Printout
0.0 g
ref 100
+
222.5 g
+ 100.00 %
Wxx%
+
222.500 g
+ 144.94 %
+
322.5 g
+ 144.94 %
%
0.00