CAL U
��
CAL O
��
30770
CAL O
��
CAL 5
��
200.0
250.0
82077
250.0
MCB_MPT-BAPS-e-1912
TARE
, the appeared triangle must be located in the
Press
upper left part of the display.
If not, press again
HOLD
Press
repeatedly until "CAL 0" appears.
TARE
Press
,
the display will show a numeric value.
ENTER
Then press
,
"CAL 0" appears.
HOLD
Press
,
"CAL 5" appears.
TARE
Press
Enter the required size of the adjustment weight (See
Section 1, "Technical Data"):
For this select the digit to be changed with
numeric value using
ENTER
Acknowledge by
Place the adjustment weight carefully in the centre of the
weighing plate; a numerical value will appear in the display.
ENTER
Press
The adjustment process is started.
The balance will automatically return to Weighing mode upon
successful completion of the adjustment and will display the
weight of the adjustment weight.
Take away adjustment weight.
TARE
.
TARE
.
.
HOLD
and the
35