The displayed weight is
permanently changing
The weighing result is
obviously incorrect
Should other error messages occur, switch balance off and then on again. If the
error message remains inform manufacturer.
14.1 Adjustment
Observe stable environmental conditions. A warming up time (see chapter 1) is
required for stabilization.
Display
z000
��
a
SETUP
��
UNKT
��
B
CAL
��
K
CAL U
��
34
 Draught/air movement
 Table/floor vibrations
 The seat surface/weighing plate is in contact with
foreign bodies or is not correctly positioned.
 Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
 The display of the balance is not at zero.
 Adjustment is no longer correct.
 Great fluctuations in temperature.
 The balance is on an uneven surface.
 Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
Operation
 Start balance by pressing
 Press
for approx. 3 sec,
in the display appears „SETUP", followed by „UNIT"
TARE
 Press
repeatedly until "CAL iB" appears
HOLD
 Press
, „CAL U" appears
ON
OFF