Télécharger Imprimer la page

Tecnimed VisioFocus Pro Mode D'emploi page 28

Publicité

Les langues disponibles

Les langues disponibles

1.
The projection/display does not turn on: the batteries are completely dead or
incorrectly inserted; replace or reinsert them (par. #11).
2.
The temperature is not projected between the two arches: the unit is not at the correct
distance; move the thermometer forward or back until the projected temperature falls
precisely between the two arches (fig. 6).
3.
The projected temperature is not clearly visible: there is too much light in the room
or the batteries are low; cast a shadow over the subject or replace the batteries.
4.
The sensor (fig. 1) is damaged or water has seeped into the thermometer: contact
your Dealer right away for technical service.
5.
The thermometer temperature reading is too low:
- make certain that the conditions outlined in the warnings (par. #2) have been met;
- check that the waveguide (fig. 1) is not soiled or damaged; if it is, clean it as
indicated in par. #12 or contact your Dealer for technical service;
- check that the thermometer is perpendicular to the forehead as indicated in fig. 3.
6.
The temperature reading is too high: make certain that the conditions outlined
DESCRIPTION: the projection/display reads "Lo.5".
PROBLEM: the room temperature is too low (below 10°C/50°F if you've pressed the
"FACE"
button or below 5°C/41°F if you've pressed the "HOME"
SOLUTION: move to another, warmer room and, if you are taking the body
temperature, wait for the stabilization of the device and of the subject.
DESCRIPTION: the projection/display reads "Hi.2".
PROBLEM: the temperature detected exceeds the limit for operation (>42.5°C /
>108.5ºF if you've pressed the "FACE"
pressed the "HOME"
SOLUTION: if you've pressed the "FACE"
pressed the correct button, with regard to the measured object, and that the warnings
have been met.
If you've pressed the "HOME"
cannot be taken (because it is too high).
DESCRIPTION: the projection/display reads "Lo.3".
PROBLEM: if you've pressed the "FACE"
to be too low (<34.0°C or <93.2°F).
If you've pressed the "HOME"
below the operating limit (<1°C/33.8°F).
SOLUTION: if you've pressed the "FACE"
soiled or damaged and that the subject has not come from a cold room.
If you've pressed the "HOME"
DESCRIPTION: the display shows "off".
MEANING: the button that you've pressed is disabled when the device is set in NURS.
SOLUTION: if you want to enable it see par. 7.
10. TROUBLESHOOTING
12 - English -
DESCRIPTION: the projection/display shows "Hi.2" and the
value alternately.
MEANING: ATTENTION! The temperature is above the level
alarm you've chosen (see par. 7).
button or >80°C / >176°F if you've
button).
button: please make certain that you've
button (DOCT mode): the temperature reading
button: the forehead temperature appears
button (DOCT mode): the surface temperature read is
button: make certain the waveguide is not
button: the temperature reading cannot be taken.
button).