• within 10 seconds focus the thermometer on an internal wall (not the inside of an
external wall) or wardrobe with uniform temperature and at a point approximately
80/150 cm (30 to 60 inches) from the floor. Press the "HOME"
button (fig. 10);
10
9
Fig.
Fig.
• once the right distance is reached (temperature value between the arches, fig. 5),
release the button: the lights flash slowly and the display shows the room
temperature.
To ensure a reliable temperature reading, do not focus the thermometer on an
outside wall, window, source of heating or cooling (radiator, air conditioner, lamp,
computer, surface in contact with the human body, etc.).
The thermometer is now ready to take a reading and will keep the MQCS for 30 minutes.
"MQCS" on the display indicates that manual quick calibration has been performed.
This system enables the thermometer to take accurate readings.
The MQCS can also be performed in case of countdown (see par. 8.2).
In "NURS" setting, the MQCS is mandatory every 30 minutes.
In "DOCT" setting, see the par. #8.2.
8.2 AQCS - countdown (only DOCT MODE)
In DOCT setting, if the thermometer is handled at length or if there is a significant
temperature difference with respect to the room temperature, a countdown will
appear on the display indicating that you should wait until automatic temperature
calibration AQCS (Automatic Quick Calibration System) has been completed.
The countdown will continue updating as long as differences in temperature are
detected (for example, because the thermometer has been continuously handled).
At this point you have 2 options:
1. make the MQCS as described in par. #8.1
or
2. without touching the thermometer, wait until the countdown has run its course
and automatic quick calibration AQCS (countdown) is completed.
At the end of the countdown, the thermometer can take accurate readings.
"AQCS" appears on the display indicating that automatic quick calibration has
been performed.
10 - English -