The current settings are highlighted by the green backlight.
• When the new desired setting appears, release the button. The purple backlight
will turn green.
Only one setting can be modified at a time.
NOTES:
- The measurement of the body temperature must be taken always at the centre
of the forehead (par. #4.1) or at the eyelid (par. #4.2), no matter what the current
setting is: in fact, the axillary, oral, rectal or core settings provide a forehead
temperature value that is a valid approximation of the axillary, oral or rectal/internal
temperature respectively (see par. #3).
- if the thermometer was produced with the sole internal temperature setting, the
ORAL, RECTAL, AXILLA and CORE settings will not appear in the sequence.
- Temperature level alarm: if you choose 37.0°C the thermometer alternates the
projection of "Hi.2" and the temperature above 37°C; if you choose 38.0°C the
thermometer alternates the projection of "Hi.2" and the temperature above 38°C.
- Air function: choose
in case of high air conditioning in the room (in case of
high air conditioning in the room, when
appears, wait for
and release
the button - the display will show alternately the word
and the temperature
value); otherwise choose
.
8. ROOM TEMPERATURE CALIBRATION
8.1 MQCS
The exclusive Manual Quick Calibration System (MQCS) promptly corrects the
device's internal temperature and adapts it to the real temperature of the room
where the reading is to be taken.
This system is extremely useful in case of prolonged use, when you need to keep
the temperature of the thermometer steady, or in case you have to move quickly
from one room to another - having different temperature - and you cannot wait too
long to have the thermometer stabilized to room temperature: for example, if you
move from room to room in a hospital department, especially if rooms have different
exposures (North, South, etc.) or in case of ambulance use for home-care.
This system is very useful also if the device is kept in the pocket or if handled for
long time for example during screening of passengers in airports.
Proceed as follows (room temperature must be in the 10-45°C/50.8-113°F
temperature range):
• press the "FACE"
and "HOME"
buttons (fig. 9) simultaneously;
• the symbol CAL
will appear and the display will light up in blue. Open the
cap;
09 - English -