4.3 Use on newborns in incubators
For use in the incubator do the MQCS inside (see par. #8):
• press simultaneously "FACE"
appears on the display;
• within 10 sec open the protective cap and, handling the device according to the
position of the forehead of the baby, introduce it in the incubator, press the
"HOME"
button and focus the lights on the inner wall of the incubator.
• Release the button.
Using the "FACE"
(fig. 8).
4.4 Other readings
In "DOCT" mode, using the "HOME" button, VisioFocus PRO can also be used
to:
1) read the temperature of objects and liquids in the 1-80°C (33.8-176°F)
temperature range. For example: canteen meal, liquids, blood bag, incubator,
professional tools, room temperature, etc.;
2) scan the skin temperature of different body areas, in order to detect possible
inflammations and/or circulatory diseases or any other problem which may cause
an alteration of the surface skin temperature. Of course VisioFocus PRO can be
used on open wounds or organs during surgical interventions, ensuring maximum
hygiene thanks to the total absence of contact.
These measurements are not entirely accurate and must be considered as
relative rather than absolute values. If taken on the skin, they can show the
difference between two close or symmetrical areas.
Proceed as you would for a forehead reading but press the "home"
the display will light up in green and show the house symbol
In "NURS" mode the "HOME" button is disabled; anyway, using the "FACE"
button, you can also take any temperature from 1 to 80°C/33.8 to 176.0°F, but
in this case, these measurements are not entirely accurate and must be
considered as relative rather than absolute values. If taken on the skin, they can
show the difference between two close or symmetrical areas. The display will
show alternately the temperature value and "Lo.3", if the recorded value is
<34°C/93.2°F or "Hi.2", if it is >37°C or 38°C (98.6 or 104°F) depending on the
Alarm level - see par. #7.