Les langues disponibles

5. ROOM TEMPERATURE
When the thermometer is in stand-by, the room temperature and the symbol
flash and remain displayed automatically for 45 minutes (DOCT mode) or 30
minutes (NURS mode) after the last reading is taken.
To display the room temperature on the backlit display, while the thermometer
is in stand-by mode, press the "Mem"
button: the display illuminates in
orange showing the room temperature and the symbol
flashing.
6. MEMORY FUNCTION
The "Mem" button (in DOCT mode or in NURS mode if it is enabled) lets you call
up the last 9 temperature readings.
Press twice the "Mem"
button: the display will light up in violet and the value
of the last reading will be shown accompanied by the number 1 and the symbol
or
depending on which button was used for that reading.
Pressing the button again calls up the second to last, third to last reading and so
on, accompanied by the numbers 2, 3, etc.
7. HOW to CHANGE the SETTINGS
Depending on where it is going to be sold, your thermometer leaves the factory:
• Celsius (
) or Fahrenheit (
) degrees;
• referred to oral (
), rectal (
) or axillary (
) or internal temperature
readings (
) or only internal temperature readings (no symbol will be
shown);
• "NURS" mode (
), (highly recommended if used by nurses, in
ambulances and for any intensive use, such as in airports etc: "HOME" and "Mem"
buttons are disabled; MQCS is requested and mandatory every 30 minutes) or
"DOCT" mode (
), (suggested for doctors' use: "HOME" and "Mem"
buttons are enabled; AQCS is automatic and MQCS is optional);
• ≥37.0°C (≥98.6°F) or ≥38.0°C (≥100.4°F) temperature level alarm: you can
choose the temperature threshold beyond which the thermometer alternates the
projection of "Hi.2" and the temperature (for 7 seconds);
• memory function enabled (
) or disabled (
);
• "air" function enabled (
) or disabled (
): to be activated
in case of high air conditioning in the room, to minimize the cooling effect of intense
air conditioning on the subject/patient.
If necessary, these settings can be modified as follows:
• while the thermometer is off or in stand-by mode, press the "Mem"
button an
hold it down; after about 8 seconds the visualization on the display changes showing
in rotation the following combinations:
08 - English -