Product Description; Fields Of Application - 3M ESPE Paradigm DeepCure Mode D'emploi

Lampe à photopolymériser led
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Light guide

Product Description

Paradigm DeepCure is a high-performance LED light
source for the polymerization of light-curing dental
materials. The device has two components: a wireless
handpiece with a built-in battery which can be replaced
by 3M ESPE customer care and a charger. The device
is a medical electrical device in accordance with
IEC 60601-1 and is available as a tabletop device.
Wall-mounting is not possible.
In comparison with conventional light-curing devices,
Paradigm DeepCure features excellent beam collimation
and a uniform beam profile, directing more of the
light energy to the restoration being polymerized and
producing a deep, uniform, and complete curing of
the restoration.
The light source is a high-performance light diode (LED).
The beam emerging from the device covers the light
wavelength range of 430 to 480 nm relevant, for
instance, for camphor quinone products and is suitable
for use with the majority of light-curing dental materials,
including materials for fillings, liners, core build-ups,
fissure sealings, temporary restorations, and cements
for indirect restorations.
See the manufacturer's information for the exposure
time required for the specific dental material.
Settable exposure times:
• 5, 10, 15, 20 sec
• Continuous mode (120 sec)
• Tack-cure mode (1 sec)
Place the handpiece on a flat surface when not in use.
The handpiece can be connected to the charger between
applications so that the battery is charged. The hand-
piece must be connected to the charger at the latest
when the battery power level display glows red steadily
Glare shield
Handpiece
so that the battery is charged (cf. "Battery Power Level
Display on Handpiece").
The device is shipped with a light guide with 10 mm
diameter. It is not permissible to use the light guides of
other devices.
The handpiece is equipped with a sleep mode to mini-
mize the device's energy consumption. The handpiece
switches to sleep mode if it is not used for a period of
about 5 minutes or the incorrect charging voltage is
detected.
The charger uses a maximum of 0.2 W when ready for
operation.
These Operating Instructions should not be
discarded for the duration of use of the device.

Fields of Application

• Polymerization of light-curing dental materials with
photo initiator for the wavelength range 430-480 nm.
- Though the majority of light-curing dental materials
are responsive in this range of wavelengths,
you may wish to contact the manufacturer of the
material in question.
Technical Data
Charger
Operating voltage:
Nominal consumption:
Dimensions without
country-specific adapter: Length: 65 mm
Weight:
Classification:
Manufacturer:
Model:
Handpiece
Power supply:
Utilizable wavelength
range:
Wavelength peak:
Light intensity (between
400 and 515 nm):
Light emission area:
100-240 V 50/60 Hz
0.2 A max
Width: 40 mm
Depth: 31 mm
75 g
Protection class II,
Click Technology Co., Ltd.
CPS 008050100
Lithium-ion battery,
nominal voltage 3.7 V,
capacity 2300 mAh
430-480 nm
444-452 nm
1470 mW/cm
2
-10%/+20%
(independent of battery power
level)
60-65 mm
2
(optically active)
3