Appendix A: EMC Information

Guidance and Manufacturer's Declaration- Electromagnetic Emissions:
This device is intended for use in the electromagnetic environment specified below. The
user of this device should make sure it is used in such an environment.
Emissions Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC61000-3-2
Voltage fluctuations
/ Flicker emissions
IEC61000-3-3
WARNING
1. The device should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is
necessary, the device should be observed to verify normal operation in the configuration in which it will be used.
2.Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this
equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this
equipment and result in improper operation.
3. Portable RF communications equipment (including peripherals such as antenna cables and external antennas)
should be used no closer than 30 cm (12 inches) to any part of the PUMP, including cables specified by the
manufacturer. Otherwise, degradation of the performance of this equipment could result.
Guidance and Manufacturer's Declaration- Electromagnetic Immunity:
This device is intended for use in the electromagnetic environment specified below. The
user of this device should make sure it is used in such an environment.
Basic EMC
Professional
Standard
Healthcare
Environment
Electrostatic
±8kV contact
Discharge (ESD)
±15kV air
IEC61000-4-2
Electrical fast
±2kV for power supply line
transient/ burst
±1kV for input/output line
IEC61000-4-4
± 1 kV line(s) to
Surge
line(s)
IEC61000-4-5
± 2 kV line(s) to earth
Compliance
Electromagnetic Environment-Guidance
The device uses RF energy only for its internal function.
Group1
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment
Class B
The device is suitable for use in all establishments, including
domestic establishments and those directly connected to the
public low-voltage power supply network
Class A
Complies
Immunity Test Levels
Home
Healthcare
Facility
Environment
± 1 kV line(s) to
line(s)
Compliance
Levels
±8kV contact
±15kV air
±2kV for power
supply line
±1kV for input/
output line
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
18
Electromagnetic
Environment-Guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Mains power quality should be that of
atypical commercial or hospital environ-
ment
Mains power quality should be that of
atypical commercial or hospital environ-
ment.