8. Troubleshooting
8.1 Twister (No. 1826)
Error
ON/OFF switch
fails to work.
Motor fails to
start.
No, or insufficient
vacuum (not in
the green range
on the scale).
Vacuum release /
venting, too slow.
Vacuum pump
doesn't start.
The vacuum pump
switches on and
off intermittently
during the mixing
process.
8.2 Twister venturi (No. 1827)
Error
ON/OFF switch
fails to work.
Motor fails to
start.
No, or insufficient
vacuum.
EN
Cause
•
Power not connected.
•
Device fuse blown.
•
Motor fault.
•
Suction filter system
plugged.
•
Seal surfaces dirty.
•
Solenoid valve fault.
•
Vacuum pump fault.
•
Suction filter system
plugged.
•
Solenoid valve fault.
•
Bowl detection pin stick-
ing.
•
At <100% vacuum, the
vacuum pump will switch
off once the set vacuum
has been reached.
Cause
•
Power not connected.
•
Device fuse blown.
•
Motor fault.
•
Compressor switched
off.
•
Operating pressure too
low or too high.
•
Compressed air hose
not connected, leaky or
bent.
•
Air hose too long.
•
Square are of air hose
too small.
•
Suction filter system
plugged.
•
Check the power supply.
•
Replace the fuse (Sec. 3.4).
•
Have the device repaired.
•
Have filter cleaned or changed
(Sec. 3.3).
•
Clean the seal surfaces (Sec. 3.2).
•
Have the device repaired.
•
Have the device repaired.
•
Have filter cleaned or changed
(Sec. 3.3).
•
Have the device repaired.
•
Clean area around pin.
•
Have the device repaired.
•
None
If the vacuum should drop due to
seepage, the vacuum pump will
automatically switch back on until
the set vacuum level has been
reached.
•
Check the power supply.
•
Replace the fuse (Sec. 3.4).
•
Have the device repaired.
•
Switch the compressor on.
•
Check dynamic operating pressure
(see technical data, Sec. G.2).
•
Switch off other consumers.
•
Check the hose.
•
Maximum length 2 m.
•
Minimum inner diameter 4 mm.
•
Have filter cleaned or changed
(Sec. 3.3).