4.
PRECAUTIONS
Important notice:
The use of apex locators alone without a preoperative and postoperative radiograph
is not recommended practice, since apex locators may not be able to work properly
in all conditions. It is recommended to take an X-ray prior to the use of the unit, and
compare the information given by both means. In case of clinically relevant discrepancy,
it is recommended to take a second radiograph with a file in place at the selected
working length [European Society of Endodontology. Quality guidelines for endodontic
treatment: consensus report of the European Society of Endodontology. International
Endodontic Journal,39, 921–930, 2006.].
It is important to follow the precautions below and pay close attention to any condition or
situation that may influence the electrical conductivity during the procedure.
Inaccurate or incorrect readings are likely to occur in the following cases, all procedure
related:
- Partially blocked canal.
- Size of the measuring file differing significantly from the canal diameter. Ideally, the
selected file should be the thickest one capable of reaching the apex.
- Presence of liquids and/or tissue debris in the access cavity. Prior to the use of
the device, the access cavity must be dried with a cotton pellet in order to prevent
leaking current.
- Contact of the file or the lip clip with metallic dental structures. Be particularly careful
with patients fitted with metal crowns or bridges.
- Contact of the file with another instrument.
- Very dry canal, for instance in the presence of restoration. In this case the canal
must be moistened with an irrigation solution, or with Glyde
- Contact between the file and the gums (this may cause a false reading indicating
that the apex has been reached).
- Use of an ultrasonic scaler with the counter electrode attached to the patient
(electrical noise from the scaler could interfere with the apex localization).
- Use of the apex locator in conjunction with an electric scalpel.
- Use of a damaged lip clip, hook or fork.
For apex localization, concentrations of NaOCl higher than 5 % may result in reduced
accuracy.
As a safety precaution in order to avoid over-instrumentation, it is recommended to
proceed as follows: place the file onto an endodontic ruler, where the propex
indicates '0.0'. Subtract a minimum of 0.5 mm from the measured file length.
Please also respect the following precautions:
- For your own safety, be aware of wearing personal protective equipment (gloves,
glasses, mask).
- If the bar graph makes sudden large movements in the coronal part of the canal,
slowly continue advancing the file toward the apex until the signal returns to normal.