TROUBLESHOOTING
Please review the checklist below should you experience a problem with your
propex
pixi
. If the problem persists after following the proposed solutions, please contact
your distributor.
Warning: The following patient related factors may prevent accurate readings:
- Blocked root canals,
- Teeth with large apices,
- Root fracture or perforation,
- Metal crowns or bridges, if they come into contact with the file or the lip clip.
#
Problem
1
During battery charging the
battery symbol blinks at a
fast rate.
2
The device does not turn
on by pressing the "ON /
OFF" button.
3
The device shuts off during
the procedure.
4
No sound during the
procedure.
Possible cause
1. The battery is not
connected.
2. The battery is not
rechargeable.
1. The button is
functioning incorrectly.
2. The battery is
discharged.
3. Electronic malfunction.
1. The battery is low.
1. The sound control is
set at "Mute" level.
21
Solution
1. Open the battery
compartment, and
connect the battery
as described in the
User Manual, section
6.4.
2. Replace the battery
with a rechargeable
type as described
in the user manual,
section 6.4.
1. Try pressing the "ON
/ OFF" button several
times. If the device
still does not turn on,
please contact your
distributor.
2. Charge the battery.
3. Contact your
distributor.
1. Charge the battery.
1. Adjust the sound
level by pressing the
"VOLUME" button.