Packet Types
Packet Type
ICP Waveform
Mean ICP Trend
System Status
ICP Waveform Packet Description
Description
Start of Packet
Product ID
Device SN
Protocol Version
Packet Type
Sensor Serial Number
Date
Time
ICP Value (1 through 100)
Data Counter
Packet Checksum
End of Packet
Content
1 second of data (100 samples)
1 minute of data (6 samples)
Physiological alarm status
Data markers
Sensor not connected
Sensor not zeroed
Fault condition
String
*
CERELINK
CLKxxxxxxx
V01
ICPWV
SNxxxxxxx
YYYY/MM/DD
HH:mm:ss
(-/+)xxx.x
00001
0xFFFF (initial value)
\r\n
Frequency
1 packet every second
1 packet every minute
1 packet every 30 seconds
Size (bytes)
1
8
10
3
5
9
10
8
6 * 100
5
4
2
39
EN - ENGLISH
Details
-
-
-
-
Waveform data
-
-
-
100 consecutive values
Increment with each message
Reset when 99999 is reached
CRC16
x^16 + x^12 + x^5 +1
-