Problem (cont.)
ICP reading goes negative
TROUBLESHOOTING
Error Codes :
Error Code
Cause
2027 2036
Battery not installed or defective.
2026
Very low battery. Not enough battery
charge to power up the system.
2024 2025 2040
Internal memory corruption
100C
System initialization test failure.
2000 2014 2039
1010 1011 1012 1013 1014
Internal components failures.
1015 1016 1017 1018
201C 202A 202F 203B
2017 2018 2019 2029
2030 2031 2033 2034
2035
1009 100A 100B 101A
Internal communication failure.
101D
2001 2002 2003 2004
2005
200F 201A 202B 202D
202E 203C
2010 2013 2015 2021
2032 2038
100D 203E 1000
Input channel error.
100E 100F 1004 1005
Output channel error.
1019
2028
Internal system temperature error.
201B 203D 2016 2037
Internal clock failure.
202C
Button stuck
Cause (cont.)
Recommendation (cont.)
System not grounded
• If problem persists, replace extension cable.
properly.
• If problem persists, replace ICP Sensor.
• If problem persists, replace power supply.
Electrical noise from
• If problem persists, contact Technical Support.
adjacent devices.
Recommendation
Replace battery.
Refer to Replacing the Battery section.
Charge the battery by connecting the monitor to the power
supply.
The monitor can be used while connected to the power supply.
Allow the unit to reboot or force a reboot by holding down the
power button.
If problem persists contact Technical Support.
Allow the unit to reboot or force a reboot by holding down the
power button. If problem persists contact Technical Support.
Allow the unit to reboot or force a reboot by holding down the
power button.
If problem persists, charge the battery for at least 30 minutes
by connecting the monitor to the power supply.
If problem persists contact Technical Support.
Allow the unit to reboot or force a reboot by holding down the
power button.
If problem persists, charge the battery for at least 30 minutes
by connecting the monitor to the power supply.
If problem persists contact Technical Support.
Allow the unit to reboot or force a reboot by holding down the
power button.
If problem persists contact Technical Support.
Allow the unit to reboot or force a reboot by holding down the
power button.
If problem persists contact Technical Support.
Allow the unit to reboot or force a reboot by holding down the
power button. If problem persists contact Technical Support.
Internal back up battery needs replacement.
Contact Technical Support.
Power on button or mute button stuck or help for more than 30
seconds. Allow the unit to reboot or force a reboot by holding
down the power button. If problem persists contact Technical
Support.
28