SETUP FOR CLINICAL USE
Return to the Codman CereLink ICP Monitor and press the arrow to continue the synchronization process (see Figure 21).
4
Figure 21: External Patient Monitor Setup – Arrow to Continue
Use the available calibration signals (0, 20 and 100mmHg) on the ICP Monitor to confirm the correct calibration and scaling of
5
the External Bedside Monitor.
Figure 22: External Patient Monitor Setup – Calibration Signal
Refer to the Instructions for Use of the External Bedside Monitor to optimize the calibration.