Operation (continued)
sensor reference point
(flange face)
Low Calibration Point
Distance from Sensor Reference Point to Low
Calibration Point: usually process empty level.
Values
Range: 0.0000 to 100.000 m
High Calibration Point
Distance from Sensor Reference Point to High
Calibration Point: usually process full level.
Values
Range: 0.0000 to 100.000 m
Wizard Complete
BACK, CANCEL, FINISH (Display returns to 1.1 Quick Start Wizard menu when Quick
Options
Start is successfully completed.)
To transfer Quick Start values to the device and return to Program menu, press
Then press
to return to Measurement mode.
SITRANS LR560 Communications: HART
• You will need the full manual to acquire the list of applicable parameters.
• We recommend that you use SIMATIC Process Device Manager (PDM) to program your
device.
• Application Guides for setting up HART devices with SIMATIC PDM can be downloaded
from the product page of our website at: www.siemens.com/LR560 under Support.
Page EN-12
SITRANS LR560 (HART) – QUICK START MANUAL
High
Calibration
Point
distance
Low
Calibration
Point
space
level
(Finish).
7ML19985XF81