Publicité

3. When the probe temperature has sta-
bilized, continue stirring the water with a
plastic straw or swizzle, while taking si-
multaneous temperature measurements
with the probe and noncontact thermom-
eter. Hold thenoncontact thermometer
within 3 inches of the surface of the water
for most accurate measurement.
4. Infrared temperature measurement
should be within ±1°C (±2°F) of probe
reading or 0°C (32°F).

5.2.2 Hot Calibration Check

Use the same procedure as cold cali-
bration, substituting hot water.
1. Fill a large styrofoam cup (10 ounce or
larger) up to 1/2 to 1 inch below the rim
with hot water (>140°F/60°C). Hot tap
water is adequate for the procedure.
2. Repeat steps 2 and 3 above.
Important: The surface of the water
must be agitated while taking the IR
temperature measurement.
5.2.2 Vérification de
l'étalonnage par le chaud
Utilisez la même procédure que pour
l'étalonnage par le froid, en replaçant
l'eau froide par de l'eau chaude.
1) Remplissez un gobelet de polystyrène
de 30 cl ou plus avec de l'eau chaude
(> 60 °C / 140 °F) jusqu'à 1 ou 2
centimètres du bord. L'eau chaude du
robinet convient pour cette procédure.
2) Répétez les étapes 2 et 3, ci-dessus.
Important : la surface de l'eau doit être
agitée
pendant
température infrarouge.
la
prise
de