Measuring Signal Output; Supply Voltage - HBM MC2A Manuel D'emploi

Amplificateur de mesure pour capteurs inductifs
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

MC2A

3.5 Measuring signal output

"5 V, R
y2,5 kΩ
L
(Option: 4 mA ... 20 mA, R
The output signal of the fully modulated measuring amplifier is
available at connection 5 and 6 (operating voltage zero). The permitted load
resistance must be
If the MC2A is equipped with a current output stage, an impressed current of
+ 4 mA (variable zero point) to 20 mA (nominal output signal).
The permissible ballast resistance
the mode of operation − voltage output or current output − is selected (circle
visible, position 1¢ voltage output).

3.6 Supply voltage

The measuring amplifier MC2A is suitable for connection to an unstabilized
DC voltage of + 10.5 V ... 26 V. The maximum current drain will be < 75mA.
For thermal dissipation reasons it is advisable to select a voltage less than
20 V. The supply voltage shall be connected to clamps No. 8 and 9. The posi-
tive pole shall be No. 9.
The amplifier is protected against wrong polarities of the supply voltage.
WARNING
The power supply leads should only be attached to points 8 and 9 other-
wise damage may be caused to the amplifier.
A1974−2.2 en/de/fr
< 400 Ω)
L
y
2.5 k
or above.
R
L
is x 400
. With the slide switch S 22,
13
"
5 V, being
HBM