4 Operating

4.1
1. Prior to putting on the belt, remove any residues of skin cream or
ointment. The skin must be grease-free. Do not place the belt on
Putting on the belt
injured or inflamed areas of skin (wounds, spots, rashes, reddening, etc.).
2. Remove the plastic films marked "on"of the electrode pads,
Handle these films carefully so that they can be stuck back in
place to protect the pads after each use.
3. Put the belt
treated. If your backache is at waist level, you will be suffering
from pain in the upper treatment area. Position the belt so that
the middle of the belt is at waist level.
4. If your back pain is in the lower treatment area, position the belt so
that the middle of the belt sits approx. 3 - 5 cm below the waist.
5. When the belt has been manoeuvred into the correct position,
bring the two ends of the neoprene straps together and push in
the clip fastener
6. The belt may be tightened or loosened by proceeding as described
in "3.3 Adjust belt width".
4.2
1. The unit is switched on by setting the intensity controllers,
Switching on
and
9
, to the desired strength. The unit offers 8 intensity settings
for treatment strength from 0 - 80 mA.
the unit
2. The unit's basic setting now appears on the display
3. Turn the two intensity controllers on the top of the TENS unit in
C
CH
CH
a clockwise direction (1 = weaker/8 = stronger) until the electric
P1
P2
P3
P4
pulses of both channels can be felt. The stimulation should be
P5
P6
P7
P8
perceived as pleasant.
15
9
CAUTION
Subjective perception of the electric current may alter with
changing frequencies, pulse widths or current shapes.
Regulate the intensity as soon as the application is
perceived as unpleasant or the pleasant prickling sensation
is not felt for a longish period.
4.3
The unit offers 8 preset treatment programs. The programs differ
in respect of varying pulse widths and frequencies.
Selecting
Program
Pulse intensity
the program
P1
P2
P3
P4
P5
P6
P7
P8
8
15:06
Pagina 5
6
and
4
on. Position it on the area of the body to be
5
to fasten the belt tight.
14
.
Pulse width
Frequency
Mode
max.
80 mA
260 μs
15 Hz
Constant
80 mA
260 μs
100 Hz
Varying
80 mA
260 μs
120 Hz
Constant
80 mA
260-150 μs
2-100 Hz
Varying
80 mA
260 μs
100 Hz
Varying
80 mA
260 μs
80<->7 Hz
Varying
80 mA
260 μs
120 Hz
Varying
80 mA
P1 - P7
Sequential
7
.
15
4.4
Selecting the
treatment time
4 Operating
1. The unit has been switched on by adjusting the intensity
controllers,
15
and
9
, to one of the 8 intensity settings (see 4.2).
2. The display
14
shows the unit's basic setting.
3. Push the keypad cover
12
downwards. Do this by pressing with
the thumb on the area on the cover identified with raised dots.
Press the MODE key
13
repeatedly until the desired treatment
program (P1 - P8) is framed on the display.
NOTES
Each of these programs starts as soon as it is selected using
the MODE key
13
regardless of whether or not a treatment
time has been set. The preset treatment time on first use is
15 minutes.
Turn the two intensity controllers,
15
and
to the OFF position to stop the program.
4. The most recently set treatment program is stored. The most recently
used program is set and flashes when the unit is switched on again.
1. The unit has been switched on and a treatment program has
been selected (see 4.2 and 4.3).
2. The display
14
shows the unit's basic setting with the program
that has been selected. When the unit is first switched on, the
display flashes for the preset treatment duration "15" or the
most recently set treatment time.
3. The unit offers 4 preset treatment times, 15 , 30 and 60 minutes,
continuous and a setting for continuous operation (©).
4. Push the keypad cover
12
downwards. Press the TIMER key
repeatedly until the desired treatment time (15 - 30 - 60 min. or
© for continuous operation) flashes on the display. Push the
keypad cover
12
back up.
NOTES
The treatment time always begins immediately as soon as
one of the three preset times has been selected or altered.
Turn the two intensity controllers,
15
and
to the OFF position to stop the program.
5. The appropriate display "15", "30" or "60" flashes for the duration
of the treatment time set. The unit switches off automatically
when the treatment time has expired. The unit does not switch off
independently in continuous mode "©".
6. The most recently set treatment time is stored. The most recently
selected treatment duration is set and flashes when the unit is
switched back on.
GB