Message / fault
SENSOR-ERROR
SENSOR-ERROR
91 - 93
NO ROTOR OR
ROTORCODE
ERROR
WRONG ROTOR !!!
N > ROTOR MAX
FC Init Error
FATAL EEPROM
1 - 5
ERROR
34

Switch on automatic circuit breaker

(only for types 1701-01 and 1706-01)
Switch off the mains switch and separate the centrifuge from the mains!
Press the plastic pin (A) on the automatic circuit breaker.
Reconnect the centrifuge to the mains supply.
35

Acceptance of the centrifuges for repair

If the centrifuge is returned to the manufacturer for repair, it must be decontaminated and cleaned to protect persons,
environment and material.
We reserve the right to accept contaminated centrifuges.
Costs incurred for cleaning and disinfection are to be charged to the customer.
We ask for your understanding in this matter.
36

Disposal

When you are disposing of the device, the respective statutory rules must be observed.
Pursuant to guideline 2002/96/EC (WEEE), all devices supplied after August 13, 2005 may not be disposed as part of
domestic waste. The device belongs to group 8 (medical devices) and is categorized in the business-to-business
field.
The icon of the crossed-out trash can shows that the device may not be disposed as part of domestic
waste.
The waste disposal guidelines of the individual EC countries might vary. If necessary, contact your
supplier.
40/105
Cause
90
Error / defect electronics
Error / defect unbalance sensor
No rotor installed.
Faulty speedometer.
(for type 1701-30 only)
A rotor is installed which isn't approved for
type 1701-30.
Speed in the selected program greater than
the maximum speed of the rotor.
Error / defect electronics
Error / defect electronics
− Perform a MAINS RESET.
− Open lid.
− Install rotor.
− Open lid.
− Install rotor 1798.
− Check the set speed.
Reduce the set speed
− Error / defect electronics
− Perform a MAINS RESET.
A