Preparation For Use

General Precautions

Make sure the unit is always grounded and secure during use. Do not disable power cord ground connection. Grounding
reliability is achieved only when power cord is connected to a hospital-grade receptacle. Inspect the electrical plug and cord
routinely. Do not use if damaged. Do not use the video colposcope in the presence of any flammable anesthetics.
Do not open the video colposcope housing. An electrical shock hazard exists due to high voltage. There are no user
serviceable parts inside the video colposcope, except the lamp.
NOTE: Opening of the video colposcope housing by an unauthorized repair facility will void the product warranty.
WARNING: Total system risk current should not exceed 50µA. An isolation transformer is required if the
total system risk current exceeds 50µA when accessories are interconnected.
CAUTION: The video colposcope is cooled via a vent fan located in the back of the unit. The fan draws
air from beneath the video colposcope and exhausts the air out the back of the video colposcope.
Verify that the unit is no less than 6" from a wall.
WARNING: DO NOT use a converter adapter that will convert the three-prong AC plug to a two-prong
line plug. The power supply in the video colposcope will not be properly grounded and electric shock
might result.
CAUTION: The colposcope can be damaged if the unit is transported while holding the handle.
The unit shjould be transported by grasping the pole.

Assembly of Video Colposcope

Before initial set up of the video colposcope, check all components received against the parts list of components (see
Components section of this manual) to verify a complete set. If parts are missing, please notify Welch Allyn. Review the
Nomenclature, Preparation for Use, Operation, and Maintenance sections to become familiar with the equipment.

Vertical Pole Stand Assembly

The video colposcope stand is shipped unassembled. Minimal assembly is required.
1. To assemble the stand, remove the stand and base parts from their cartons.
2. Place base upside down on the floor and insert casters into holes on the bottom of the base. (Do not place two locking
casters next to each other.)
3. Place the base on the floor with casters down. Lock the two locking casters.
Locking caster
12
Locking caster