Summary of Warnings and Cautions

WARNING: All video monitors and peripheral equipment used with this instrument must be IEC 601-1
(or UL 2601) listed to comply with medical electrical safety guidelines. If not IEC 601-1(UL-2601) listed,
the monitor/peripheral equipment should be positioned outside of the 6-foot patient contact area.
Please contact Welch Allyn for more information.
WARNING: Total system risk current should not exceed 50µA. An isolation transformer is required if the
total system risk current exceeds 50µA when accessories are interconnected.
CAUTION: The video colposcope is cooled via a vent fan located in the back of the unit. The fan draws
air from the front of and beneath the video colposcope and exhausts the air out the back of the video
colposcope. Verify that the unit is no less than 6" from a wall.
WARNING: DO NOT use a converter adapter that will convert the three-prong AC plug to a two-prong
line plug. The power supply in the video colposcope will not be properly grounded and electric shock
might result.
CAUTION: DO NOT clean illumination lens with alcohol.
DO NOT touch optical or illumination lenses except as described in Maintenance section of this manual.
WARNING: The lamp operates at a high temperature. DO NOT attempt to remove the lamp before
allowing it to cool. Allow at least five minutes for the lamp to cool before replacing. Replace with
Welch Allyn lamp #09800 only.
CAUTION: The colposcope can be damaged if the unit is transported while holding the handle.
The unit should be transported by grasping the pole.