Daily
Weekly
Lupo
20
RotoForce-1
Instruction Manual
5. Maintenance
Clean all accessible surfaces of the machine with a moist cloth
To avoid clogging of the Lupo nozzle, you should clean the nozzle
once a week:
n Set the mode switch on Lupo to STOP
n Remove Lupo from the supporting arm.
n Hold the Lupo lubricant bottle firmly and remove the top lid.
n Empty the bottle, still mounted on Lupo, of any remaining
lubricant.
−
Fill the bottle with mild, soapy water.
−
Put Lupo back on the supporting arm.
−
Set the mode switch on Lupo to START
−
Set the dosing control to max. dosing (step 10) and let the
water run through the nozzle for 2-3-min.
−
Repeat the above points but fill with tap water to rinse
out any soap residue.
−
Refill with lubricant and remount the top lid.
−
Clean the O-rings and lubricate them using High
Vacuum Grease
o
position.
o
.