2) Dry the orthotic joints with a cloth or allow them to air dry. Avoid expos­
ure to direct heat (e.g. from an oven or radiator).
7 Maintenance
INFORMATION
The product may be exposed to increased loads by the patient.
► Shorten the maintenance intervals according to the expected loads.
The manufacturer requires at least a semi-annual inspection of the product
to verify functionality and check for wear.
Spare parts are listed in the section "Components/design".
Only use 633F7 special lubricant.
17K29: Replacing the ring lock
1) Loosen the joint screw and remove the bearing nut.
2) Disassemble the orthotic joint.
3) Remove the clamping sleeve and disassemble the ring lock.
4) Replace the ring lock and assemble the clamping sleeve.
5) Assemble the orthotic joint with the screw and the bearing nut.
6) For final assembly, secure the joint screw with Loctite 241.
8 Disposal
Dispose of the product in accordance with national regulations.
9 Legal information
All legal conditions are subject to the respective national laws of the country
of use and may vary accordingly.
9.1 Liability
The manufacturer will only assume liability if the product is used in accord­
ance with the descriptions and instructions provided in this document. The
manufacturer will not assume liability for damage caused by disregarding the
information in this document, particularly due to improper use or unauthor­
ised modification of the product.
9.2 CE conformity
The product meets the requirements of Regulation (EU) 2017/745 on medic­
al devices. The CE declaration of conformity can be downloaded from the
manufacturer's website.