CAUTION
Excessive strain due to use on more than one patient
Risk of injury and loss of functionality as well as damage to the product
► Use the product on only one patient.
► Observe the maintenance recommendations.
NOTICE
Thermal overloading of the product
Damage due to improper thermal treatment
► Do not carry out any heat treatment at temperatures above 300 °C
(570 °F).
► Prior to thermal treatment, remove all temperature-critical components
(such as plastic parts).
NOTICE
Exposure of the product to unsuitable environmental conditions
Damage, brittleness or destruction due to improper handling
► Avoid storage in condensing ambient humidity.
► Avoid contact with abrasive substances (e.g. sand, dust).
► Do not expose the product to temperatures below -10 °C (14 °F) or
above +60 °C (140 °F) (e.g. sauna, excessive sunlight, drying on a
radiator).
INFORMATION
Increased wear and tear
The orthotic knee joints are intended only for use in combination with free-
motion or foot lifting orthotic ankle joints. Wear occurs more rapidly in
combinations involving orthotic ankle joints with a dorsal stop.
5 Preparing the product for use
INFORMATION
Parallel alignment of the orthotic joints
Use the 743R6 alignment fixture to position the orthotic joints on the plaster
positive.