5.1 Alignment
INFORMATION
There is a plastic protector on the product's pyramid. It protects the con­
necting section from scratches during the alignment and fitting of the pros­
thesis.
► Remove the protector before the patient leaves the workshop/fitting
area.
5.1.1 Applying/removing the footshell
Recommended tools: 2C100 changing device
>
1) Slide the prosthetic foot into the footshell.
2) Press the heel of the prosthetic foot into the footshell until it engages.
Recommended tools: 2C100 changing device
>
1) Push the lock of the footshell to the rear and pull up the heel of the pros­
thetic foot.
2) Remove the prosthetic foot from the footshell.
5.1.2 Bench Alignment
TT bench alignment
Required materials: 662M4 goniometer, 743S12 heel height measuring apparatus,
743A80 50:50 gauge, alignment apparatus (e.g. 743L200 L.A.S.A.R. Assembly or
743A200 PROS.A. Assembly)
Perform the assembly and alignment of the prosthetic components in the alignment
apparatus according to the following specifications:
Sagittal plane
Heel height: Effective heel height (shoe heel height - sole thickness of fore­
   
foot) + 5 mm
Exterior foot rotation: approx. 5°
   
   
Anterior placement, middle of the prosthetic foot to the alignment reference
line: 30 mm
   
Connect the prosthetic foot and prosthetic socket using the chosen adapters.
The instructions for use of the adapters must be observed.
   
Determine the centre of the prosthetic socket with the 50:50 gauge. Align the
prosthetic socket centrally to the alignment reference line.
Socket flexion: Individual residual limb flexion + 5°
Frontal plane
Alignment reference line of prosthetic foot: between big toe and second
   
toe
14
Bench alignment process

Publicité