Ottobock 1C40 C-Walk Instructions D'utilisation page 11

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 18
1.2 Combination possibilities
This prosthetic component is compatible with Ottobock's system of modular
connectors. Functionality with components of other manufacturers that have
compatible modular connectors has not been tested.
2 Intended use
2.1 Indications for use
The product is intended exclusively for lower limb exoprosthetic fittings.
2.2 Area of application
Our components perform optimally when paired with appropriate compon­
ents based upon weight and mobility grades identifiable by our MOBIS clas­
sification information and which have appropriate modular connectors.
The product is recommended for mobility grade 3 (unrestricted
outdoor walker) and mobility grade 4 (unrestricted outdoor walk­
kg
er with particularly high demands).
Size [cm]
Max. body weight [kg]
2.3 Environmental conditions
Storage and transport
Temperature range –20 °C to +60 °C (–4 °F to +140 °F), relative humidity 20 % to
90 %, no mechanical vibrations or impacts
Allowable environmental conditions
Temperature range: –10 °C to +45 °C (14 °F to 113 °F)
Moisture: relative humidity: 20% to 90%, non-condensing
Chemicals/liquids: fresh water as dripping water, occasional contact with salty air
(e.g. near the ocean)
Solids: dust
Prohibited environmental conditions
Chemicals/moisture: salt water, perspiration, urine, acids, soapsuds, chlorine water
Solids: dust in high concentrations (e.g. construction site), sand, highly hygroscopic
particles (e.g. talcum)
24 to 25
75
26 to 30
100
11

Publicité

Table des Matières
loading

Table des Matières