Chapter 4: Control Testing; Why Perform Control Tests; About The Control Solutions - Roche ACCU-CHEK Aviva Manuel De L'utilisateur

Masquer les pouces Voir aussi pour ACCU-CHEK Aviva:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Chapter 4: Control Testing

Why Perform Control Tests

Performing a control test lets you know that your meter and test strips are working properly to give
reliable results. You should perform a control test when:
• You open a new box of test strips
• You left the test strip container open
• You want to check the meter and test strips
• Your test strips were stored in extreme
temperatures

About the Control Solutions

• Only use Accu-Chek Aviva Control Solutions.
• Your meter automatically recognizes the
control solution.
• Write the date you opened the bottle on the
bottle label. The solution is good for three
months from that date or until the expiry date
on the bottle label, whichever comes first.
• You dropped the meter
• Your test result does not agree with how
you feel
• You want to check if you are testing
correctly
• Do not use control solution that is past the
expiry date.
• The solution can stain clothing. If you spill it,
wash your clothes with soap and water.
• Close the bottle tightly after each use.
• Store the bottle at 2 to 32°C. Do not freeze.
43

Publicité

Chapitres

Table des Matières
loading