Controls And Indicators; Scope Of Delivery - VOLTCRAFT PH-212 Notice D'emploi

Soil ph measuring device set
Table des Matières

Publicité

Measuring devices and their accessories are not toys and should be
kept away from children!
In commercial institutions, the accident prevention regulations of
the relevant professional insurance association for electrical sys-
tems and operating materials are to be observed.
In schools, training centres, amateur and self-help workshops, han-
dling of measuring instruments must be supervised by trained per-
sonnel in a responsible manner.
Do not switch on the measuring instrument immediately after it has
been taken from a cold to a warm environment. The condensated
generated as a result could spoil the device.
reach room temperature before switching it on.
The supplied durable pH electrode is suitable for soft-muddy media
such as planting soils, liquids etc. Stones and other hard foreign
particles in the measuring media can damage the electrode. Any
application of a deliberate mechanical load, will make the warranty
null and void.

Scope of delivery

PH-212 pH measuring device
Durable pH electrode with Epoxy casing and storage solution
pH4 and pH7 buffer/calibration solutions
Operating instructions
Storage case
16
Allow the device to

Controls and indicators

1. LC display
6. Calibration trimmer pH7
2. ON/OFF button
7
3. HOLD button
8. BNC connecting plug
4. BNC connection-
9. electrode grip
electrode socket
10. electrode storage solution
5. battery compartment
Commissioning
Inserting the batteries
Before the commissioning of this meter, you must first install a new
9 V battery pack. Insertion of battery is described in the „Mainte-
nance and Cleaning" section.
Calibration
Calibration is necessary in order to optimally adapt the electrode to
the pH meter. An optimal pH electrode outputs a voltage of 0 mV at
a temperature of 25°C and a pH value of 7.00. However, since each
electrode is different and deviates from the optimum depending
upon the temperature and the wear and tear, it must always be
adapted to the pH measuring device for achieving the highest pos-
sible accuracy.
Calibration trimmer pH4
17

Publicité

Chapitres

Table des Matières
loading

Ce manuel est également adapté pour:

10 11 22

Table des Matières