GeoMax Zeta125 Manuel De L'utilisateur page 51

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

SERVICE menu
Menu
Step
Description
3.
Use the arrow up and down keys on the remote control to change the
calibration value. The height of the laser beam changes with the cali-
bration value.
4.
Wait until SET is displayed.
5.
Repeat step 3. and 4. if necessary.
6.
To save the position:
Press the lock symbol on the remote control.
To leave the calibration mode without storing the change:
Turn off the laser.
The SERVICE menu is PIN protected. The menu is only accessible by authorised
service centres.
Next step
Step Description
1.
Move the cursor to EXIT.
2.
Press the Enter key
to return to the SET UP menu.
Zeta125/ Zeta125s | 50