Rugby 260SG / 270SG / 280DG, Checking Level Accuracy

Checking Level Accuracy

It is the responsibility of the user to follow operating instructions and to period-
ically check the accuracy of the instrument and work as it progresses.
The Rugby is adjusted to the defined accuracy specification at the factory. It is
recommended to check your Rugby for accuracy upon receipt and periodically
thereafter to ensure accuracy is maintained. If your Rugby requires adjustment,
contact your nearest authorized service center or adjust the Rugby using the
following procedure.
Do not enter this mode or attempt adjustment unless you plan to change the
accuracy. Accuracy adjustment should be performed only by a qualified individual
that understands basic adjustment principles.

Procedure

To check level accuracy of your Rugby, place the Rugby on a flat, level surface or
tripod approximately 30 m (100 ft) from a wall.