Monitoring The Volume Of Gas Used; Monitoring The Volume Of Gas Available - SOPRO COMEG 640-45L Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

I. MONITORING THE VOLUME OF GAS USED
The volume of gas used is shown in litres on the LCD display to the nearest deciliter
This meter is started by pressing on the button [F] and only resets to zero once the device is powered off.
If necessary set it to zero via the menu.
Use only with medical grade C02 for coelioscopy complying with European standards.
j. MONITORING THE VOLUME OF GAS AVAILABLE
1. With gas cylinder
As soon as the cylinder pressure reaches 40 bars a updated cylinder icon is displayed,
At 20 bars, a yellow light alarm light [B] and an alarm message "CYLINDER PRESSURE 20 bar" will be displayed.
From 10 bars, it is no longer possible to start the device (RUN/STOP key disabled); the cylinder must be changed.
Always have a new cylinder in the OR room.
12
0 1 1 3 8 7 C • 1 0 / 2 0 1 1
E N G L I S H
7
Flow
CO
L/min
P
Volume
P Cons.
L
mmHg
CYLINDER LOW 20 bar
20
bar
Flow
CO
L/min
P
Volume
P Cons.
L
mmHg
mmHg
2
is flickering.
mmHg
2

Publicité