Problem/possible cause
Displays error message:
H5i flip lid is not closed.
Air tubing is not connected properly.
The following message is displayed on the LCD after you try to update settings or copy data to
Card error, please remove SD card and contact service provider
the SD card:
SD card is not inserted correctly.
You may have removed the SD card
before settings were copied to the
device.
The following message is NOT displayed on the LCD after you try to update the settings using
Settings updated successfully, press any key
the SD card:
The settings were not updated.
Water splashing on your face from the H5i
The water tub is overfilled.
Condensation is forming in the air
tube and mask.
Leaking water tub
The water tub may be damaged or
cracked.
The cleanable water tub is not
assembled correctly.
Air feels too warm/cold in the mask
The temperature of the ClimateLine or
MAX
ClimateLine
air tubing is set too
high/low.
6
Solution
Please close H5i flip lid, attach tube and press any key
Close the flip lid ensuring that it clicks into place.
Connect firmly at both ends.
Ensure that the SD card is inserted correctly.
Reinsert the SD card and wait for the Home screen or the
Settings updated successfully, press any key"
"
to appear on the LCD.
Note: This message only appears once. If you re-insert the
SD card after you have updated your settings, the message
will not be re-displayed.
Contact your clinician/service provider immediately.
Check that the water level is below the maximum water
level mark.
Turn the H5i or relative humidity setting down via the menu.
Contact your service provider for a replacement.
Check for damage and reassemble the cleanable water tub
correctly.
Turn up/down the heated tubing temperature via the menu.
.