Problem/possible cause
The S9 and H5i are not connected
correctly.
Insufficient air delivered from the S9 device
Ramp time is in use.
Air filter is dirty.
Air tubing is not connected properly.
Air tubing is blocked, pinched or
punctured.
Mask and headgear are not positioned
correctly.
Incorrect air tubing selected.
The H5i flip lid is not latched correctly. Close the flip lid ensuring that it clicks into place.
The H5i flip lid seal is not fitted
correctly.
Device does not start when you breathe into the mask
Breath is not deep enough to trigger
SmartStart/Stop.
There is excessive leak.
Note: If Leak Alert is enabled, an
audible alert is activated and a high
leak message is displayed.
SmartStart/Stop is disabled.
Note: SmartStart/Stop is not available
if Leak Alert is enabled.
Solution
Ensure that the S9 and H5i are securely attached.
Wait for air pressure to build up or change ramp time.
Replace air filter.
Check air tubing.
Unblock or free the air tubing. Check the air tubing for
punctures.
Adjust position of mask and headgear.
If you are using the SlimLine, Standard or 3m air tubing
ensure that you have the correct air tubing selected via the
menu.
Make sure the flip lid seal is facing the right way up and
fitted securely.
Take a deep breath in and out through the mask.
Adjust position of mask and headgear.
Air tubing not connected properly. Connect firmly at both
ends.
Enable SmartStart/Stop.