TECHNICAL DESCRIPTION

01
02
EN
03
04
05
06
07
08
09
10
11
12
13
14
15
16
50
Channel
Intensity
Pulse amplitude
Output voltage
Power source
Size
Weight
Pulse rate
Pulse width
ON time
OFF time
Ramp time
Mode
TENS Burst mode B
TENS Normal mode
N
TENS Modulation
mode M
4 channels – with adjustable intensity
Adjustable, 0 – 100 mA (in steps of 1 mA)
with 500 ohm load on each channel
Asymmetrical, Bi-phasic square pulse
0 – 50 V
integrated 3.7V recharchable battery
13.8cm (L) x 7.8cm (B) x 2.8cm (H)
242 grams
Adjustable, from 1 – 150 Hz, 1 Hz/step
Adjustable, from 30 – 400 microseconds,
10 μ/step
Adjustable, from 2 – 90 seconds, 1 Sec./step
Adjustable, from 0 – 90 seconds, 1 Sec./step
Adjustable, from 1 – 8 seconds, 1 Sec./step
5 TENS modes: B (Burst) N (Normal),
M (Modulation), SD1 (40% modulation),
SD2 (70% modulation) – 3 EMS modes: C
(Constant), 5 (Synchronous), A (Alternate),
P (Program-Mode) – 37 TENS programs
(set by default)
Burst rate: Adjustable, 0.5 – 5 Hz – Pulse
Width adjustable, 30 – 400 μs – Frequen-
cy fixed = 100 Hz
The pulse rate and pulse width are adjus-
table. It generates continuous stimulation
based on the setting value
Modulation mode is a combination of pul-
se rate and pulse width modulation. The
pulse rate and width are automatically
varied in a cycle pattern. The pulse width
is decreased by 50% from its original
setting in 0.5 second, then the pulse rate
is decreased by 50% from its original
setting in 0.5 second. Total cycle time is
1second. In this mode, pulse rate (1 – 150
Hz) and pulse width (30 – 400 μs) are
fully adjustable.