ResMed VPAP Malibu Guide Utilisateur page 27

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

268106r1.book Page 22 Thursday, November 30, 2006 2:35 PM
CPAP mode
Mode/Ramping Indicator
CPAP pressure
C:RAMP
A>
CPAP
A>>>
Pressure Graph
Apnea Indicator
Minute ventilation (L/min)
Breath/phase/expiration
Average inspiration time
Oxygen saturation (%)
Note: (Set) values are those selected by the clinician.
The treatment screens contain the following information:
Treatment mode: Mode of treatment set by your clinician. Options are: CPAP,
S and VSet.
Ramping indicator: (CPAP mode, S mode) Appears if the VPAP Malibu is in
ramp mode. This disappears once the ramp time has elapsed.
Settling indicator: (VSet mode) Appears if the VPAP Malibu is in settling mode.
This disappears once the settling time has elapsed.
Set pressure(s): In CPAP mode, this displays the set treatment pressure
(centimeters of water). In S mode, it displays exhalation and inhalation pressures
(centimeters of water). In VSet mode, it displays the AutoSet pressure (mean
mid-point pressure between inhalation and exhalation).
Trigger indicator: (S mode) How the VPAP Malibu changes the pressure when
you are inhaling. "S" (Spontaneous) indicates a patient triggered change.
Apnea indicator: (CPAP mode, VSet mode) Visual indicator that displays an "A"
to indicate an apnea is occurring.
Pressure bar graph: Graphical display of the changing pressure.
Leak: Current mask leak (liters per minute).
Respiratory rate: Number of breaths per minute.
22
S mode
(First screens only display if Ramp or Settling are selected)
Mode/Ramping Indicator
(Set) EPAP pressure
8.0
S:RAMP
8.0-11.0
exit
S>>>>
PS 3.0
Mode
15.0
SPONT:
8.0-11.0
exit
S>>>>
PS 3.0
Trigger Indicator
Leak
LK: 24L/min
MV: 12.0
S**TiMn
TiMx 4.0
Ti 2.1s
(last 5 breaths)
SpO2 : 12%
Heart rate (BPM)
HR : 57
VSet mode
Mode/Settling Indicator
(Set) IPAP pressure
VS:SETL
A>>>>
Pressure Support
VSET
A>>>>
Apnea Indicator
Respiratory rate (bpm)
RR: 12
VT:1000
Tidal volume (mL)
(Top row: VSet and S mode only)
Ti Max setting (sec)
1: 1.6
Average I:E ratio (last 5 breaths)
(This screen only displays if ResLink
exit
and an oximeter are connected)
EPAP
IPAP
8.0-11.0
PS 3.0
(Set) PS
8.0-11.0
PS 3.0

Publicité

Table des Matières
loading

Table des Matières